Qualification test of a MPPC-based PET module for future MRI-PET scanners
- 1. Research Institute for Science and Engineering, Waseda University, 3-4-1, Ohkubo, Shinjuku, Tokyo (Japan)
- 2. Department of Radiological and Medical Laboratory Sciences, Nagoya University Graduate School of Medicine, 65-banchi, Tsurumai-cho, Showa-ku, Nagoya-shi, Aichi (Japan)
Description
We have developed a high-resolution, compact Positron Emission Tomography (PET) module for future use in MRI-PET scanners. The module consists of large-area, 4×4 ch MPPC arrays (Hamamatsu S11827-3344MG) optically coupled with Ce:LYSO scintillators fabricated into 12×12 matrices of 1×1 mm2 pixels. At this stage, a pair of module and coincidence circuits was assembled into an experimental prototype gantry arranged in a ring of 90 mm in diameter to form the MPPC-based PET system. The PET detector ring was then positioned around the RF coil of the 4.7 T MRI system. We took an image of a point 22Na source under fast spin echo (FSE) and gradient echo (GE), in order to measure interference between the MPPC-based PET and the MRI. We only found a slight degradation in the spatial resolution of the PET image from 1.63 to 1.70 mm (FWHM; x-direction), or 1.48–1.55 mm (FWHM; y-direction) when operating with the MRI, while the signal-to-noise ratio (SNR) of the MRI image was only degraded by 5%. These results encouraged us to develop a more advanced version of the MRI-PET gantry with eight MPPC-based PET modules, whose detailed design and first qualification test are also presented in this paper. - Highlights: • We developed a high resolution, compact PET module for future MRI-PET. • We found slight degradation in the resolution of PET images operating with the MRI. • The SNR of the MR images was only degraded by 5% operating with PET. • PET and MRI have little impact on each other. • We are now developing a more advanced version of the MRI-PET gantry
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2014.04.040Additional details
Identifiers
- DOI
- 10.1016/j.nima.2014.04.040;
- PII
- S0168-9002(14)00445-8;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 765
- Journal Page Range
- p. 275-279
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 9. international ''Hiroshima'' symposium on development and application of semiconductor tracking detectors
- Acronym
- HSTD-9 2013
- Dates
- 1-5 Sep 2013
- Place
- Hiroshima (Japan)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46125275
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COINCIDENCE CIRCUITS; COMPACTS; DESIGN; IMAGES; NMR IMAGING; PHOSPHORS; PHOTONS; POSITRON COMPUTED TOMOGRAPHY; RINGS; SIGNAL-TO-NOISE RATIO; SODIUM 22; SPATIAL RESOLUTION; SPIN ECHO
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; BOSONS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; ELECTRONIC CIRCUITS; ELEMENTARY PARTICLES; EMISSION COMPUTED TOMOGRAPHY; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LIGHT NUCLEI; MASSLESS PARTICLES; NANOSECONDS LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; RADIOISOTOPES; RESOLUTION; SODIUM ISOTOPES; TOMOGRAPHY; YEARS LIVING RADIOISOTOPES
Optional Information
- Copyright
- Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.