Published November 21, 2014 | Version v1
Journal article

Qualification test of a MPPC-based PET module for future MRI-PET scanners

  • 1. Research Institute for Science and Engineering, Waseda University, 3-4-1, Ohkubo, Shinjuku, Tokyo (Japan)
  • 2. Department of Radiological and Medical Laboratory Sciences, Nagoya University Graduate School of Medicine, 65-banchi, Tsurumai-cho, Showa-ku, Nagoya-shi, Aichi (Japan)

Description

We have developed a high-resolution, compact Positron Emission Tomography (PET) module for future use in MRI-PET scanners. The module consists of large-area, 4×4 ch MPPC arrays (Hamamatsu S11827-3344MG) optically coupled with Ce:LYSO scintillators fabricated into 12×12 matrices of 1×1 mm2 pixels. At this stage, a pair of module and coincidence circuits was assembled into an experimental prototype gantry arranged in a ring of 90 mm in diameter to form the MPPC-based PET system. The PET detector ring was then positioned around the RF coil of the 4.7 T MRI system. We took an image of a point 22Na source under fast spin echo (FSE) and gradient echo (GE), in order to measure interference between the MPPC-based PET and the MRI. We only found a slight degradation in the spatial resolution of the PET image from 1.63 to 1.70 mm (FWHM; x-direction), or 1.48–1.55 mm (FWHM; y-direction) when operating with the MRI, while the signal-to-noise ratio (SNR) of the MRI image was only degraded by 5%. These results encouraged us to develop a more advanced version of the MRI-PET gantry with eight MPPC-based PET modules, whose detailed design and first qualification test are also presented in this paper. - Highlights: • We developed a high resolution, compact PET module for future MRI-PET. • We found slight degradation in the resolution of PET images operating with the MRI. • The SNR of the MR images was only degraded by 5% operating with PET. • PET and MRI have little impact on each other. • We are now developing a more advanced version of the MRI-PET gantry

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2014.04.040

Additional details

Identifiers

DOI
10.1016/j.nima.2014.04.040;
PII
S0168-9002(14)00445-8;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
765
Journal Page Range
p. 275-279
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
9. international ''Hiroshima'' symposium on development and application of semiconductor tracking detectors
Acronym
HSTD-9 2013
Dates
1-5 Sep 2013
Place
Hiroshima (Japan)

Optional Information

Copyright
Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.