Simple Monte-Carlo method to calibrate well-type HPGe detectors
Creators
- 1. University Institute for Applied Radiation Physics, Grand-Pre 1, CH-1007 Lausanne (Switzerland)
- 2. Laboratorium Spiez, Bundesamt fuer Bevoelkerungschutz des VBS, CH-3700 Spiez (Switzerland)
Description
The activity concentration of radionuclides in the environment is routinely measured using axial high-purity germanium (HPGe) detectors. For low-activity samples, well-type detectors are often chosen for their high efficiency. These detectors are usually calibrated with specific radionuclides that do not allow a generalization to other sources. While Monte-Carlo methods have been used for calibration purposes, they usually need extensive and precise geometrical characterizations. The goal of this paper is to propose a simple calibration method for a well-type HPGe detector combining Monte-Carlo calculations with relatively undemanding experimental measurements. After a systematic study of the relative importance of the different geometrical parameters, the uncertainties on the crystal dimensions and especially the external dead layer thickness revealed themselves to be the factors with the largest effect on the efficiency. We therefore proposed to estimate the dead layer thickness by experimentally measuring an 152Eu source and adjusting the thickness with the computed ratio of two distinct full-energy peaks. The method has been finally validated using two traceable sources of 131I and 60Co, and the agreement was found better than 6%. This simple Monte-Carlo calibration procedure can now be applied for any source geometry of γ-emitting radionuclides having a known decay scheme
Additional details
Identifiers
- DOI
- 10.1016/j.nima.2006.09.040;
- PII
- S0168-9002(06)01658-5;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 569
- Journal Issue
- 3
- Journal Page Range
- p. 790-795
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38038261
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; COBALT 60; CRYSTALS; DECAY; EFFICIENCY; EUROPIUM 152; GAMMA SPECTROSCOPY; GEOMETRY; GERMANIUM; HIGH-PURITY GE DETECTORS; IODINE 131; LAYERS; MONTE CARLO METHOD; THICKNESS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; CALCULATION METHODS; COBALT ISOTOPES; DAYS LIVING RADIOISOTOPES; DIMENSIONS; ELECTRON CAPTURE RADIOISOTOPES; ELEMENTS; EUROPIUM ISOTOPES; GE SEMICONDUCTOR DETECTORS; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IODINE ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MATHEMATICS; MEASURING INSTRUMENTS; METALS; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; RARE EARTH NUCLEI; SEMICONDUCTOR DETECTORS; SPECTROSCOPY; YEARS LIVING RADIOISOTOPES