Published 1996 | Version v1
Book

Temperatures stabilization of a field instrument for uranium enrichment measurements

  • 1. Canberra Industries, Inc., Meriden, CT (United States)

Description

Enrichment measurements with sodium iodide (NaI) detectors are hampered with a number of problems related to the temperature behavior of NaI crystals and the associated electronics. This problem is of particular concern in applications requiring the use of fixed regions of interest; such applications are used by the International Atomic Energy Agency (IAEA) in Vienna. The Canberra IMCA is a new portable instrument for such applications which can use either a NaI or a Ge detector. In developing the IMCA to meet the IAEA requirements for NaI detectors, Canberra has designed a system with a new temperature stabilization method capable of maintaining the detector stability at 0.5% over a temperature range of -10 to +50 degrees C. This paper includes a detailed description of this IMCA temperature stabilization system, as well as test results for a range of temperatures using uranium standards

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (United States)
Imprint Title
1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
Imprint Pagination
2138 p.
Journal Page Range
p. 955-958.

Conference

Title
Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1996.
Place
Anaheim, CA (United States).

Optional Information

Secondary number(s)
CONF-961123--.