A multichannel single-photon sensitive detector for high-energy physics: the megapixel EBCCD

Description

We present a study of the characteristics of a hybrid image intensifier tube based on a thinned backside Electron Bombarded CCD (EBCCD) with better performance in spatial resolution, single photoelectron detection and gain stability than the conventional intensified systems based on a Micro Channel Plate followed by a CCD. Single photon detection sensitivity has been studied and a procedure to correct the gain non-uniformity of the EBCCD has been developed. The EBCCD, operating at 14 kV, has an average gain of ∼3000 electrons/photoelectron and a noise of the order of 100 electron/pixel. These characteristics make the EBCCD a very attractive device for many applications in high-energy physics, astrophysics and biomedicine. A possible application in a neutrino oscillation experiment is presented

Additional details

Identifiers

PII
S0168900299012656;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
442
Journal Issue
1-3
Journal Page Range
p. 154-158
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2000 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.