Published December 2008 | Version v1
Journal article

Combining prompt gamma activation analysis and off-line counting

  • 1. Hungarian Academy of Sciences, Budapest (Hungary). Institute of Isotopes
  • 2. Hungarian Academy of Sciences, Budapest (Hungary). Atomic Energy Research Institute

Description

The off-line γ-counting of in-beam activated radionuclides has been explored to extend the detection capabilities of prompt gamma activation analysis (PGAA). Such combination of the prompt measurement with a subsequent decay-counting is feasible if radionuclides with half-lives of minutes or hours are produced in the sample during neutron irradiation. Thanks to the simpler spectrum and the higher counting efficiency of decay counting, both selectivity and sensitivity can be improved. The pros and cons of the proposed method have been demonstrated on a series of industry-related measurements. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
278
Journal Issue
3
Journal Page Range
p. 657-660
ISSN
0236-5731
CODEN
JRNCDM

INIS

Optional Information

Notes
13 refs.