Published September 2010 | Version v1
Journal article

A scintillator detector for low energy projectile detection in radiative recombination and dielectron recombination experiments

  • 1. Graduate University of Chinese Academy of Sciences, Beijing (China)
  • 2. Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou (China)

Description

A new CsI (Tl)scintillation detector readout with R7525 PMT (Hamamatsu)is designed to detect low energy (<4 MeV/u) ions in radiative recombination and dielectronic recombination experiments at the main cooler storage ring in Lanzhou. The structure of the scintillator detector is described in this paper. The detector is tested with α source and ion beams, respectively. The signals from background, 200 MeV/u C6+ and alpha radioactive source are recorded and analyzed. The results show that the detector has good response to high-and low-energy ions. The maximum counting rate of the detector can reach 106 ions/s and the vacuum level near the detector can reach 10-10 Pa, both are good for recombination experiments. The installation of the new detector to CSRm is of great help for future radiative recombination and dielectronic recombination experiments. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Physics Review
Journal Volume
27
Journal Issue
3
Journal Page Range
p. 296-299
ISSN
1007-4627

Optional Information

Notes
7 figs., 1 tab., 5 refs.