Published March 15, 2014 | Version v1
Journal article

Dosimetry of an In-Line Kilovoltage Imaging System and Implementation in Treatment Planning

  • 1. Department of Radiation Oncology, Saarland University Medical Center, Homburg/Saar (Germany)
  • 2. Hochschule für Technik und Wirtschaft des Saarlandes, Saarbrücken (Germany)
  • 3. Department of Radiation Oncology, University of Minnesota, Minneapolis, Minnesota (United States)

Description

Purpose: To present the beam properties of the Siemens 70-kV and 121-kV linear accelerator-mounted imaging modalities and commissioning of the 121-kV beam in the Philips Pinnacle treatment planning system (TPS); measurements in an Alderson phantom were performed for verification of the model and to estimate the cone-beam CT (CBCT) imaging dose in the head and neck, thorax, and pelvis. Methods and Materials: The beam profiles and depth–dose curve were measured in an acrylic phantom using thermoluminescent dosimeters and a soft x-ray ionization chamber. Measurements were imported into the TPS, modeled, and verified by phantom measurements. Results: Modeling of the profiles and the depth–dose curve can be achieved with good quality. Comparison with the measurements in the Alderson phantom is generally good; only very close to bony structures is the dose underestimated by the TPS. For a 200° arc CBCT of the head and neck, a maximum dose of 7 mGy is measured; the thorax and pelvis 360° CBCTs give doses of 4-10 mGy and 7-15 mGy, respectively. Conclusions: Dosimetric characteristics of the Siemens kVision imaging modalities are presented and modeled in the Pinnacle TPS. Thermoluminescent dosimeter measurements in the Alderson phantom agree well with the calculated TPS dose, validating the model and providing an estimate of the imaging dose for different protocols

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ijrobp.2013.12.007

Additional details

Identifiers

DOI
10.1016/j.ijrobp.2013.12.007;
PII
S0360-3016(13)03630-4;

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
88
Journal Issue
4
Journal Page Range
p. 913-919
ISSN
0360-3016
CODEN
IOBPD3

Optional Information

Copyright
Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.