Published 1998 | Version v1
Report Open

Calibration of extremity dosemeters for gamma radiation fields

Description

In this work the kerma conversion factor are free in air, dose equivalent H(d,0) are presented, they were obtained theoretical and experimentally in finger and arm for gamma radiation fields. Extremity dosemeters put on surface finger and arm phantom have been irradiated. The finger phantom is a solid cylinder of PMMA polymethylmethacrylate 19 mm diameter and 300 mm height. The arm phantom is a 73 mm external diameter cylinder with PMMA walls 2.5 mm thick fill with water and 300 mm height. The radiation sources were cobalt 60 and cesium 137 from the Regional Center of Reference (CRR) of the National Commission of Atomic Energy (CNEA) and the Nuclear Regulatory Authority (ARN). Also in ISO wide X ray spectra W60, W110 and W200 have been irradiated. The results obtained show a good correlation with those published, they have a difference less than 7%. The factors will be applied to the evaluation of the equivalent doses coming from workers whose main irradiated zone is in the hands. (author)

Availability note (English)

Available from INIS in electronic form

Files

31001543.pdf

Files (85.9 kB)

Name Size Download all
md5:6117b44b02548487fc5b4736603fea30
85.9 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Calibracion de dosimetros personales para extremidades en campos de radiacion gamma

Publishing Information

Imprint Pagination
6 p.
Report number
ARN-PI--32/98

Conference

Title
4. Regional congress on radiological and nuclear safety
Original Conference Title
4. Congreso regional en seguridad radiologica y nuclear
Dates
19-23 Oct 1998
Place
La Habana (Cuba)

Optional Information

Notes
6 refs., 4 tabs.