Published 2002 | Version v1
Journal article

Monte Carlo calculation and experimental verification of the photon energy response of tooth enamel in a head-sized plexiglas phantom

Description

The use of electron paramagnetic resonance (EPR) tooth dosimetry for calculation of organ doses requires conversion of the measured absorbed dose in enamel. Before deriving conversion factors from simulation calculations with a realistic anthropomorphic human phantom, in the current study a simplified phantom was chosen to compare EPR measurement and Monte Carlo calculation. The dose response of tooth enamel of molars at various positions inside a cylindrical Plexiglas phantom of head-size was calculated by Monte Carlo modelling in parallel photon beams of X rays of 63 keV equivalent energy and 60Co gamma rays (1.25 Mev). For X ray exposure, preliminary results of EPR dosimetry with tooth enamel samples prepared from molars irradiated in the phantom were in agreement with calculation. The mean value of the ratio of the measured to the calculated dose was 0.93±0.08. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
101
Journal Issue
1-4
Journal Page Range
p. 549-552
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
13. international conference on solid state dosimetry
Dates
9-13 Jul 2001
Place
Athens (Greece)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
33067664
Subject category
S07: ISOTOPES AND RADIATION SOURCES; S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CHEMICAL DOSEMETERS; COMPUTERIZED SIMULATION; ELECTRON SPIN RESONANCE; GAMMA DOSIMETRY; MONTE CARLO METHOD; PHANTOMS; VALIDATION; X-RAY DOSIMETRY
Descriptors DEC
CALCULATION METHODS; DOSEMETERS; DOSIMETRY; MAGNETIC RESONANCE; MEASURING INSTRUMENTS; MOCKUP; RESONANCE; SIMULATION; STRUCTURAL MODELS; TESTING