Selector-integrator - device for spectrometric measurements under high counting rates
Description
A selector-integrator with the input counting rate up to 10sup(7 pulse)/s is described. The device consists of the base line restorer, the linear gates, the logic circuits, the integrator and the pile-up rejector. The selector-integrator is made of the CAMAC standard. The device principal circuits are shown and its parameters are listed. The integration time is in the range from 50nc up to 2 mc. The proposed device provides the accuracy of spectrometric measurements about 1%
Availability note (English)
MF available from INIS under the Report Number.Files
9395932.pdf
Files
(275.8 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:9db09015cd2381f52aedc6821c8fda3e
|
275.8 kB | Preview Download |
System files
(32.6 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Селектор-интегратор-прибор для спектрометрических измерений при высокой скорости счета
Publishing Information
- Imprint Pagination
- 14 p.
- Report number
- JINR-R--13-10110
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 9395932
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; BACKGROUND NOISE; CAMAC SYSTEM; COUNTING RATES; FLOWSHEETS; GATING CIRCUITS; LOGIC CIRCUITS; PERFORMANCE; PULSE INTEGRATORS; SI SEMICONDUCTOR DETECTORS
- Descriptors DEC
- DIAGRAMS; ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; INFORMATION; MEASURING INSTRUMENTS; NOISE; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS
Optional Information
- Notes
- 4 refs.; 9 figs.