Testing stand of plastic scintillation detectors of neutrons is described. Detectors are irradiated by the beam of 14 MeV neutrons produced in d + t → n + 4He reaction using the ING-07 neutron generator. The neutrons obtained pass through the forming shielding collimator made of borated polyethylene. Formation of narrow collimated beam allows one to study the spatial resolution of the neutron detectors (possibility of determination of the point of interaction of neutron with the detector substance) that is necessary for optimization of time resolution of detectors. The stand also allows to measure the absolute efficiency of detectors of various lengths at neutron energy of 14 MeV
Additional details
Additional titles
- Original title (Russian)
- Стенд для испытания сцинтилляционных детекторов нейтронов
Publishing Information
- Journal Title
- Yaderna Fyizika ta Energetika
- Journal Issue
- no.4/22
- Journal Page Range
- p. 100-103
- ISSN
- 1818-331X
INIS
- Country of Publication
- Ukraine
- Country of Input or Organization
- Ukraine
- INIS RN
- 39033181
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COLLIMATORS; EFFICIENCY; MEV RANGE 10-100; NEUTRON BEAMS; NEUTRON DETECTORS; PERFORMANCE TESTING; PLASTIC SCINTILLATION DETECTORS; SPATIAL RESOLUTION; SPECIFICATIONS; TIME RESOLUTION; TIME-OF-FLIGHT SPECTROMETERS
- Descriptors DEC
- BEAMS; ENERGY RANGE; MEASURING INSTRUMENTS; MEV RANGE; NUCLEON BEAMS; PARTICLE BEAMS; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SPECTROMETERS; TESTING; TIMING PROPERTIES