Published April 2017 | Version v1
Journal article

Type testing of model 7200 automatic TLD reader

  • 1. Department of Physics, Faculty of Basic Sciences, Islamic Azad University, Central Tehran Branch, Tehran (Iran, Islamic Republic of)
  • 2. Iran Nuclear Regulatory Authority, Tehran (Iran, Islamic Republic of)
  • 3. Radiation Application Research School, Nuclear Science and Technology Research Institute, AEOI, Tehran (Iran, Islamic Republic of)

Description

The type testing of measuring devices is one of the most important parts of a quality management system in a personal dosimetry services program. In this study, based upon the International Electrotechnical Commission (IEC) 62387 criteria, a reader-testing program was performed for a home-made personal thermoluminescent dosimetry (TLD) reader. The stability of the reader, the effects of light exposure, temperature and fluctuations of primary power supply on TLD read-outs as the main parameters were investigated in this program. Moreover, this study assesses some important criteria of dosimetry system including the non-linearity of response, reusability, after effect and overload that may include significant contribution in the performance of a reader. The results showed that the TLD reader met all requirements of the IEC for the reader tests by a large margin. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncw100

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
174
Journal Issue
1
Journal Page Range
p. 68-73
ISSN
0144-8420

Optional Information

Notes
11 refs.