Published July 1, 1995 | Version v1
Journal article

ExB ion mass spectroscopy in magnetised plasmas

  • 1. Royal Inst. of Technol., Stockholm (Sweden). Alfven Lab.
  • 2. University of Tromsoe, N-9037 Tromsoe (Norway)

Description

A spectrometer based on the ExB drift during the transit time of the ion through a cross field region has been tested in a hydrogen plasma in the Blaamann toroidal plasma device [T. Brundtland, Vacuum 43 (1992) 185]. The magnetic field B, is the field of the device. The electric field E, which is imposed and oriented perpendicular to the magnetic field, is swept with a time long compared to the ion transit time. The ions are accelerated along the magnetic field as they enters the cross field region giving them a velocity and a transit time proportional to the charge over the mass. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
361
Journal Issue
1-2
Journal Page Range
p. 253-256.
ISSN
0168-9002
CODEN