Published 1984
| Version v1
Report
Restricted
Improving the performance of the MWPC X-ray imaging detector by means of the Multi-Step Avalanche technique
Description
X-ray imaging systems based on conventional MWPC technology and artificial delay line readout techniques have been developed at RAL for several applications over a period of some eight years. It is perceived that very limited scope exists for the further improvement of the imaging capability of the standard MWPC design. Attention has therefore been turned to the possibility of exploiting the Multi-Step Avalanche (MSA) system of electron multiplication in this context. Results from a prototype system are presented which show spatial resolution better than that achieved in the MWPC systems. The facility for controlling the effective depth of the detector electronically is also demonstrated. (author)
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Publishing Information
- Imprint Pagination
- 25 p.
- Report number
- RAL--84-089
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16035218
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DIAGRAMS; ELECTRONS; IMAGES; MULTIWIRE PROPORTIONAL CHAMBER; OPERATION; PERFORMANCE; RADIATION DETECTORS; RESOLUTION; SPECIFICATIONS; X RADIATION; X-RAY RADIOGRAPHY
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; ELEMENTARY PARTICLES; FERMIONS; INFORMATION; IONIZING RADIATIONS; LEPTONS; MEASURING INSTRUMENTS; PROPORTIONAL COUNTERS; RADIATIONS