Published 1988 | Version v1
Book

MTF-insensitive algorithm for absolute diameter measurements of coronary arteries

Description

The authors are investigating a new algorithm for the determination of coronary artery dimensions that has a low sensitivity to the modulation transfer function of the imaging system. This algorithm uses the first zero crossing frequency of the Fourier transform of the vessel profile to indicate the vessel size. Measurements of vessel phantoms imaged with blurring aperture sizes extending from 20% to 80% of the vessel diameter have shown a 10%-25% improvement in measurement accuracy over standard videodensitometric and edge detection techniques. Preliminary in vivo measurements of stenosed coronary arteries in a dog model indicate a similar improvement in accuracy

Additional details

Publishing Information

Publisher
Radiological Society of North America Inc.
Imprint Place
Oak Brook, IL (USA)
Imprint Title
Radiological Society of North America 74th scientific assembly and annual meeting (Abstracts)
Imprint Pagination
395 p.
Journal Page Range
p. 369.

Conference

Title
74. scientific assembly and annual meeting of the Radiological Society of North America (RSNA).
Dates
27 Nov - 2 Dec 1988.
Place
Chicago, IL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
21023971
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; BIOLOGICAL MODELS; CORONARIES; DOGS; FEASIBILITY STUDIES; FOURIER TRANSFORMATION; IMAGE PROCESSING; IN VIVO; PHANTOMS
Descriptors DEC
ANIMALS; ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; INTEGRAL TRANSFORMATIONS; MAMMALS; MOCKUP; ORGANS; STRUCTURAL MODELS; TRANSFORMATIONS; VERTEBRATES

Optional Information

Secondary number(s)
CONF-8811134--.