Published 1988
| Version v1
Book
MTF-insensitive algorithm for absolute diameter measurements of coronary arteries
Description
The authors are investigating a new algorithm for the determination of coronary artery dimensions that has a low sensitivity to the modulation transfer function of the imaging system. This algorithm uses the first zero crossing frequency of the Fourier transform of the vessel profile to indicate the vessel size. Measurements of vessel phantoms imaged with blurring aperture sizes extending from 20% to 80% of the vessel diameter have shown a 10%-25% improvement in measurement accuracy over standard videodensitometric and edge detection techniques. Preliminary in vivo measurements of stenosed coronary arteries in a dog model indicate a similar improvement in accuracy
Additional details
Publishing Information
- Publisher
- Radiological Society of North America Inc.
- Imprint Place
- Oak Brook, IL (USA)
- Imprint Title
- Radiological Society of North America 74th scientific assembly and annual meeting (Abstracts)
- Imprint Pagination
- 395 p.
- Journal Page Range
- p. 369.
Conference
- Title
- 74. scientific assembly and annual meeting of the Radiological Society of North America (RSNA).
- Dates
- 27 Nov - 2 Dec 1988.
- Place
- Chicago, IL (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 21023971
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; BIOLOGICAL MODELS; CORONARIES; DOGS; FEASIBILITY STUDIES; FOURIER TRANSFORMATION; IMAGE PROCESSING; IN VIVO; PHANTOMS
- Descriptors DEC
- ANIMALS; ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; INTEGRAL TRANSFORMATIONS; MAMMALS; MOCKUP; ORGANS; STRUCTURAL MODELS; TRANSFORMATIONS; VERTEBRATES
Optional Information
- Secondary number(s)
- CONF-8811134--.