Determination of energy windows for the triple energy window scatter correction method in I-131 on a Siemens SYMBIA gamma camera: a GATE simulation study
- 1. Molecular Imaging Research Center, Korea Institute Radiological and Medical Sciences, 75 Nowon-gil, Nowon-gu, Seoul 139-706 (Korea, Republic of)
- 2. Department of Radiation Convergence Engineering, and Research Institute of Health Science, Yonsei University, 1 Yonseidae-gil, Wonju,Gangwon 220-710 (Korea, Republic of)
Description
Image correction for scattered photons is important for the quantification of gamma-camera imaging using I-131. Many previous studies have addressed this issue but none have compared scattered photon fractions of I-131 with varying energy windows, to determine optimal main- and sub-energy windows for the implementation of TEW correction in I-131 imaging. We assessed the scattered photon fractions and determined the optimal main- and sub- energy windows for TEW in I-131 using a Siemens SYMBIA T2 SPECT/CT using a Monte Carlo method (GATE simulation). To validate the GATE simulation code, we compared the spatial resolutions obtained experimentally and from GATE simulation, for I-123 and Tc-99m. A high-energy general purpose (HE) collimator was used to assess the scattered photon fractions measured with the I-131 radioisotope placed at eight different field-of-view locations in a water phantom (diameter 16 cm, length 32 cm), and at the center in air. To implement the TEW (triple energy window) method, two different main-energy window widths (15 and 20%) and two different sub-energy window widths (3 and 5 keV) were used. The experimental measurement and simulation results exhibited a similar pattern with < 15% difference in spatial resolution with increasing distance. The I-131 scatter fraction with 15% of the main-energy window and 5 keV sub-energy windows was similar to the ''goldstandard'' scatter fraction. Main- and sub-energy window selection for the TEW correction in I-131 is important to avoid over- or under-correction in the scatter fraction. A 15% of main energy window with 5 keV sub-energy windows were found to be optimal for implementation of the TEW method in I-131. This result provides the optimal energy window for I-131 scintigraphy data and will aid the quantification of I-131 imaging
Availability note (English)
Available from http://dx.doi.org/10.1088/1748-0221/10/01/P01004Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Instrumentation
- Journal Volume
- 10
- Journal Issue
- 01
- Journal Page Range
- p. P01004
- ISSN
- 1748-0221
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46042339
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CORRECTIONS; GAMMA CAMERAS; IMPLEMENTATION; IODINE 123; IODINE 131; SCINTISCANNING; SIMULATION; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION; TECHNETIUM 99
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CAMERAS; COMPUTERIZED TOMOGRAPHY; COUNTING TECHNIQUES; DAYS LIVING RADIOISOTOPES; DIAGNOSTIC TECHNIQUES; ELECTRON CAPTURE RADIOISOTOPES; EMISSION COMPUTED TOMOGRAPHY; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IODINE ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; NUCLEI; ODD-EVEN NUCLEI; RADIOISOTOPE SCANNING; RADIOISOTOPES; RESOLUTION; TECHNETIUM ISOTOPES; TOMOGRAPHY; YEARS LIVING RADIOISOTOPES