Published 1996
| Version v1
Journal article
A radon exhalation monitor
Description
A radon exhalation monitor has been constructed and calibrated. The monitor is based on the adsorption of radon by active charcoal and the subsequent measurement of its γ activity by means of high resolution γ spectrometry. Exhalation rates may be determined with an accuracy of better than 20% within 48 h after exposure. The detection limit is estimated on the basis of counting statistics to about 0.1 mBq.m-2s-1. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 63
- Journal Issue
- 1
- Journal Page Range
- p. 69-72.
- ISSN
- 0144-8420
- CODEN
- RPDODE
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 27048442
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S54: ENVIRONMENTAL SCIENCES;
- Descriptors DEI
- ACTIVATED CARBON; ADSORPTION; CALIBRATION; DESIGN; EMANOMETERS; FEASIBILITY STUDIES; PERFORMANCE; RADON 222
- Descriptors DEC
- ADSORBENTS; ALPHA DECAY RADIOISOTOPES; CARBON; DAYS LIVING RADIOISOTOPES; ELEMENTS; EVEN-EVEN NUCLEI; HEAVY NUCLEI; ISOTOPES; MEASURING INSTRUMENTS; NONMETALS; NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; RADON ISOTOPES; SORPTION