Published August 1994 | Version v1
Journal article

Post injection transmission scanning in a volume imaging PET camera

  • 1. Univ. of Pennsylvania, Philadelphia, PA (United States). Dept. of Radiology
  • 2. UGM Medical Systems, Philadelphia, PA (United States)

Description

Quantitative, low noise, measured attenuation correction has been established for the PENN PET 240H Volume Imaging Camera. This is achieved, without septa, using a narrow energy and sinogram position gate to minimize scatter contamination. Twelve minute transmission acquisitions with a 0.5 mCi rod source are adequate for this purpose. Post injection transmission scans (with emission activity in the FOV) suffer from emission contamination simulating transmitted gamma ray flux. This emission contamination may be measured by performing a transmission acquisition with normal transmission energy and position gating but without a transmission source. This contamination is then subtracted from the measured post injection transmission scan. Emission activity within the FOV adds to detector deadtime, whereas only a small fraction of events are accepted into the transmission position gate, resulting in a net loss of scan statistics when compared to pre-injection transmission scanning. Removal of emission contamination and compensation for this excess deadtime results in corrected attenuation coefficients close to pre-injection values. For residual activity levels typical of FDG whole-body cancer and cardiac studies the post injection measurements are within 4% of pre-injection values

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
41
Journal Issue
4PT1
Journal Page Range
p. 1526-1531.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
nuclear science symposium and medical imaging conference.
Acronym
NSS-MIC '93
Dates
30 Oct - 6 Nov 1993.
Place
San Francisco, CA (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
26024848
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
IMAGE PROCESSING; OPTIMIZATION; PERFORMANCE; POSITRON CAMERAS; POSITRON COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION
Descriptors DEC
CAMERAS; COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; GAMMA CAMERAS; RESOLUTION; TOMOGRAPHY

Optional Information

Secondary number(s)
CONF-931051--.