Waveform integration method for measuring position resolution of PET detectors
- 1. School of Nuclear Science and Technology, Lanzhou University, Lanzhou (China)
Description
The waveform integration method is used to acquire the signal. The 2-D position map of the PET detector is obtained, and the look-up table is calculated using the watershed algorithm. A 10 × 10 LYSO with a pixel size of 2.0 mm and 1.6 mm and a 1.6 mm 12 × 12 crystal array detector are directly coupled to the photomultiplier tube. Based on the 1 GHz sampling frequency ADQ synchronous data acquisition card, the optimal PET detector 2-D position map was obtained by waveform integration method. Experiments show that 10 × 10 LYSO crystals with 2.0 mm, 1.6 mm pixels, and 12 × 12 PET detectors with 1.6 mm, the positional discrimination of individual crystal strips reaches 0.28 mm, 0.20 mm, and 0.19 mm F WHM, the average peak-to-valley ratio is 3.23, 4.26, and 3.46, respectively. The LYSO array crystal PET detector module can be used in the construction of PET system. . (authors)
Additional details
Publishing Information
- Journal Title
- Nuclear Electronics and Detection Technology
- Journal Volume
- 40
- Journal Issue
- 2
- Journal Page Range
- p. 346-352
- ISSN
- 0258-0934
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 55086785
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ALGORITHMS; CRYSTALS; DATA ACQUISITION; GHZ RANGE; MAPS; PEAKS; PHOTOMULTIPLIERS; POSITRON COMPUTED TOMOGRAPHY; RESOLUTION; SAMPLING; SIGNALS; WAVE FORMS
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DATA PROCESSING; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; FREQUENCY RANGE; MATHEMATICAL LOGIC; PHOTOTUBES; PROCESSING; TOMOGRAPHY
Optional Information
- Notes