Measured Neutron Spectra and Dose Equivalents From a Mevion Single-Room, Passively Scattered Proton System Used for Craniospinal Irradiation
Creators
- 1. Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas (United States)
- 2. Department of Nuclear Engineering, Idaho State University, Pocatello, Idaho (United States)
- 3. Department of Radiation Oncology, Washington University, St. Louis, Missouri (United States)
Description
Purpose: To measure, in the setting of typical passively scattered proton craniospinal irradiation (CSI) treatment, the secondary neutron spectra, and use these spectra to calculate dose equivalents for both internal and external neutrons delivered via a Mevion single-room compact proton system. Methods and Materials: Secondary neutron spectra were measured using extended-range Bonner spheres for whole brain, upper spine, and lower spine proton fields. The detector used can discriminate neutrons over the entire range of the energy spectrum encountered in proton therapy. To separately assess internally and externally generated neutrons, each of the fields was delivered with and without a phantom. Average neutron energy, total neutron fluence, and ambient dose equivalent [H* (10)] were calculated for each spectrum. Neutron dose equivalents as a function of depth were estimated by applying published neutron depth–dose data to in-air H* (10) values. Results: For CSI fields, neutron spectra were similar, with a high-energy direct neutron peak, an evaporation peak, a thermal peak, and an intermediate continuum between the evaporation and thermal peaks. Neutrons in the evaporation peak made the largest contribution to dose equivalent. Internal neutrons had a very low to negligible contribution to dose equivalent compared with external neutrons, largely attributed to the measurement location being far outside the primary proton beam. Average energies ranged from 8.6 to 14.5 MeV, whereas fluences ranged from 6.91 × 106 to 1.04 × 107 n/cm2/Gy, and H* (10) ranged from 2.27 to 3.92 mSv/Gy. Conclusions: For CSI treatments delivered with a Mevion single-gantry proton therapy system, we found measured neutron dose was consistent with dose equivalents reported for CSI with other proton beamlines.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ijrobp.2015.12.356Additional details
Identifiers
- DOI
- 10.1016/j.ijrobp.2015.12.356;
- PII
- S0360-3016(15)27194-5;
Publishing Information
- Journal Title
- International Journal of Radiation Oncology, Biology and Physics
- Journal Volume
- 95
- Journal Issue
- 1
- Journal Page Range
- p. 249-257
- ISSN
- 0360-3016
- CODEN
- IOBPD3
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48097233
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- DEPTH DOSE DISTRIBUTIONS; DOSE EQUIVALENTS; ENERGY SPECTRA; EVAPORATION; GY RANGE; IRRADIATION; MEV RANGE 10-100; NEUTRON BEAMS; NEUTRON FLUENCE; NEUTRON SPECTRA; PEAKS; PROTON BEAMS; THERMAL NEUTRONS
- Descriptors DEC
- ABSORBED DOSE RANGE; BARYONS; BEAMS; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; HADRONS; MEV RANGE; NEUTRONS; NUCLEON BEAMS; NUCLEONS; PARTICLE BEAMS; PHASE TRANSFORMATIONS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSE RANGES; SPATIAL DOSE DISTRIBUTIONS; SPECTRA
Optional Information
- Copyright
- Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.