Published April 1993 | Version v1
Journal article

Serial measurement of relative changes in net magnetization, 1

  • 1. Emory Univ., Atlanta, GA (United States). School of Medicine

Description

We assessed to what extent relative changes in net magnetization could be measured with the low field (0.043 T) MR imager. By the procedure to stabilize the whole MR imager hardware, we could measure the minute relative changes of the net magnetization (less than 1%) from the forearm presumably related to blood volume change. This method may be useful to measure physiological changes of blood volume in various human tissues though we need further development of hardware to measure minute relative changes in MR signal intensity with small volume selection. (author)

Additional details

Additional titles

Subtitle (English)
Assessment of its sensitivity to detect brain function using a low field MR imager

Publishing Information

Journal Title
Nippon Jiki Kyomei Igakkai Zasshi
Journal Volume
13
Journal Issue
2
Journal Page Range
p. 127-131.
ISSN
0914-9457
CODEN
NJKZEV