Published March 21, 2008 | Version v1
Journal article

Measurement of decay time of liquid scintillator

  • 1. Key Laboratory of Particle Astrophysics, Institute of High Energy Physics, Beijing 100049 (China)
  • 2. Experimental Center, Institute of High Energy Physics, Beijing 100049 (China)

Description

In this paper we describe a single photon method to measure the decay time of two kinds of homemade liquid scintillator. We find that the fast and slow decay time constants of these two samples, with 5 cm diameter and 4 cm height, are about ∼4ns and ∼17ns, respectively. These results are checked through a reference measurement by recording the scintillation pulse shapes at the output of a fast response PMT directly by a digital oscilloscope. The time responses of these two samples are as good as those of commercial liquid scintillator

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2008.01.077

Additional details

Identifiers

DOI
10.1016/j.nima.2008.01.077;
PII
S0168-9002(08)00097-1;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
587
Journal Issue
2-3
Journal Page Range
p. 300-303
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
39066299
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
HEIGHT; LIQUID SCINTILLATORS; NEUTRINO OSCILLATION; PHOTONS; PULSE SHAPERS; SCINTILLATIONS
Descriptors DEC
BOSONS; DIMENSIONS; ELECTRONIC CIRCUITS; ELEMENTARY PARTICLES; MASSLESS PARTICLES; PHOSPHORS; PULSE CIRCUITS; SIGNAL CONDITIONERS

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.