Published April 15, 1985 | Version v1
Journal article

Silicon strip detectors with capacitive charge division

  • 1. Deutsches Elektronen-Synchrotron (DESY), Hamburg (Germany, F.R.)
  • 2. Hamburg Univ. (Germany, F.R.). 2. Inst. fuer Experimentalphysik
  • 3. Politecnico di Milano (Italy). Ist. di Fisica
  • 4. Max-Planck-Institut fuer Physik und Astrophysik, Muenchen (Germany, F.R.)
  • 5. Technische Univ. Muenchen (Germany, F.R.)

Description

Measurements of the behaviour of various high-resolution silicon strip detectors with capacitive charge division have been made in a high-energy particle beam. The results are compared with an electrostatic model which calculates the charges deposited on the strips after the passage of a minimum ionizing particle. Good agreement is found and the model is used to propose a method of improving the charge-division properties of such detectors. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
235
Journal Issue
3
Series
CODEN: NIMRD.;Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
481-487
ISSN
0168-9002

Optional Information

Notes
Also published as report MPI-PAE/Exp.El.--142.