Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the VSL, Netherlands and the BIPM in mammography x-rays
- 1. Bureau International des Poids et Mesures, Pavillon de Breteuil, F-92312 Sevres Cedex, (France)
- 2. SL, P O Box 654, 2600 AR Delft, (Netherlands)
Description
A first key comparison has been made between the air-kerma standards of the VSL and the BIPM in mammography x-ray beams. The results show the standards to be in agreement within the standard uncertainty of 5.7 parts in 103. The results are analysed and presented in terms of degrees of equivalence for entry in the BIPM key comparison database. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1088/0026-1394/51/1A/06025Additional details
Identifiers
Publishing Information
- Journal Title
- Metrologia
- Journal Volume
- 51
- Journal Issue
- no.1A
- Journal Page Range
- p. 1-13
- ISSN
- 0026-1394
- CODEN
- MTRGAU
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 46045570
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AIR; CALIBRATION STANDARDS; CORRECTIONS; DATA COVARIANCES; INTERLABORATORY COMPARISONS; IONIZATION CHAMBERS; KERMA; X RADIATION
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; FLUIDS; GASES; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; STANDARDS