Independent dose calculation in IMRT for the Tps Iplan using the Clarkson modified integral
Creators
- 1. Instituto Privado de Radioterapia, Obispo Oro 423, X5000BFI Cordoba (Argentina)
Description
Intensity-Modulated Radiation Therapy (IMRT) treatments require a quality assurance (Q A) specific patient before delivery. These controls include the experimental verification in dose phantom of the total plan as well as dose distributions. The use of independent dose calculation (IDC) is used in 3D-Crt treatments; however its application in IMRT requires the implementation of an algorithm that allows considering a non-uniform intensity beam. The purpose of this work was to develop IDC software in IMRT with MLC using the algorithm proposed by Kung (Kung et al. 2000). The software was done using Matlab programming. The Clarkson modified integral was implemented on each flowing, applying concentric rings for the dose determination. From the integral of each field was calculated the dose anywhere. One time finished a planning; all data are exported to a phantom where a Q A plan is generated. On this is calculated the half dose in a representative volume of the ionization chamber and the dose at the center of it. Until now 230 IMRT planning were analyzed carried out ??in the treatment planning system (Tps) Iplan. For each one of them Q A plan was generated, were calculated and compared calculated dose with the Tps, IDC system and measurement with ionization chamber. The average difference between measured and calculated dose with the IDC system was 0.4% ± 2.2% [-6.8%, 6.4%]. The difference between the measured and the calculated doses by the pencil-beam algorithm (Pb) of Tps was 2.6% ± 1.41% [-2.0%, 5.6%] and with the Monte Carlo algorithm was 0.4% ± 1.5% [-4.9%, 3.7%]. The differences of the carried out software are comparable to the obtained with the ionization chamber and Tps in Monte Carlo mode. (author)
Files
45110592.pdf
Files
(406.4 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:ec37621007696d40ff7cf3243b26882e
|
406.4 kB | Preview Download |
Additional details
Additional titles
- Original title (Spanish)
- Calculo independiente de dosis en IMRT para el TPS Iplan usando la integral modificada de Clarkson
Publishing Information
- Publisher
- Sociedad Mexicana de Irradiacion y Dosimetria
- Imprint Place
- Mexico, D. F. (Mexico)
- Imprint Pagination
- 11 p.
- Report number
- INIS-MX--2802
Conference
- Title
- 14. International Symposium on Solid State Dosimetry
- Original Conference Title
- 14. Conferencia Internacional sobre Dosimetria de Estado Solido
- Dates
- 13-16 Apr 2014
- Place
- Cusco (Peru)
INIS
- Country of Publication
- Mexico
- Country of Input or Organization
- Mexico
- INIS RN
- 45110592
- Subject category
- S97: MATHEMATICAL METHODS AND COMPUTING;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALGORITHMS; BEAMS; COMPARATIVE EVALUATIONS; COMPUTER CODES; IMPLEMENTATION; INTEGRALS; IONIZATION CHAMBERS; MONTE CARLO METHOD; PATIENTS; PHANTOMS; PLANNING; PROGRAMMING; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; RINGS; VERIFICATION
- Descriptors DEC
- CALCULATION METHODS; DOSES; EVALUATION; MATHEMATICAL LOGIC; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOLOGY; STRUCTURAL MODELS; THERAPY
Optional Information
- Funding organization
- Organizacion Panamericana de la Salud, Washington D. C. (United States); Universidad Nacional de San Antonio de Abad del Cusco, Cusco (Peru); Instituto Peruano de Energia Nuclear, Lima (Peru); Elekta AB (Sweden); RadPro International GmbH (Germany); ThermoFisher Scientific, Inc. (United States); Varian Medical Systems (United States)