Published August 2014 | Version v1
Miscellaneous Open

Independent dose calculation in IMRT for the Tps Iplan using the Clarkson modified integral

Description

Intensity-Modulated Radiation Therapy (IMRT) treatments require a quality assurance (Q A) specific patient before delivery. These controls include the experimental verification in dose phantom of the total plan as well as dose distributions. The use of independent dose calculation (IDC) is used in 3D-Crt treatments; however its application in IMRT requires the implementation of an algorithm that allows considering a non-uniform intensity beam. The purpose of this work was to develop IDC software in IMRT with MLC using the algorithm proposed by Kung (Kung et al. 2000). The software was done using Matlab programming. The Clarkson modified integral was implemented on each flowing, applying concentric rings for the dose determination. From the integral of each field was calculated the dose anywhere. One time finished a planning; all data are exported to a phantom where a Q A plan is generated. On this is calculated the half dose in a representative volume of the ionization chamber and the dose at the center of it. Until now 230 IMRT planning were analyzed carried out ??in the treatment planning system (Tps) Iplan. For each one of them Q A plan was generated, were calculated and compared calculated dose with the Tps, IDC system and measurement with ionization chamber. The average difference between measured and calculated dose with the IDC system was 0.4% ± 2.2% [-6.8%, 6.4%]. The difference between the measured and the calculated doses by the pencil-beam algorithm (Pb) of Tps was 2.6% ± 1.41% [-2.0%, 5.6%] and with the Monte Carlo algorithm was 0.4% ± 1.5% [-4.9%, 3.7%]. The differences of the carried out software are comparable to the obtained with the ionization chamber and Tps in Monte Carlo mode. (author)

Files

45110592.pdf

Files (406.4 kB)

Name Size Download all
md5:ec37621007696d40ff7cf3243b26882e
406.4 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Calculo independiente de dosis en IMRT para el TPS Iplan usando la integral modificada de Clarkson

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico, D. F. (Mexico)
Imprint Pagination
11 p.
Report number
INIS-MX--2802

Conference

Title
14. International Symposium on Solid State Dosimetry
Original Conference Title
14. Conferencia Internacional sobre Dosimetria de Estado Solido
Dates
13-16 Apr 2014
Place
Cusco (Peru)