SU-F-T-606: Monte Carlo Evaluation of Tissue Inhomogeneity Corrections in the Treatment of Liver Cancer Patients Using Stereotactic Body Radiotherapy

  • 1. University of Kansas Hospital, Kansas City, KS (United States)

Description

Purpose: To evaluate the dosimetric performance of X-ray Voxel Monte Carlo(XVMC) algorithm in effort to clinically validate Monte Carlo-approach in heterogeneous liver phantom and liver SBRT plans. Methods: An anthropomorphic RPC liver phantom incorporating a liver structure with two cylindrical targets and organs-at-risk (OARs) was used in phantom validation. Subsequently, five patients with metastatic liver cancer were treated using heterogeneity-corrected pencil-beam(PB-hete)algorithm were analyzed following RTOG-1112 criteria .ITV was delineated on MinIP and OARs were contoured on MeanIP images of 4D-CT. PTV was generated from ITV with 5-10mm uniform margin. Mean PTV was 81.3±46.4cc. Prescription was 30–45Gy in 5 fractions, with at least PTV(D95%)=100%. Hybrid SBRT plans were generated with noncoplanar/3D-conformal arcs plus static-beams at Novalis-TX consisting of HD-MLC and 6MV-SRS. SBRT plans were re-computed using XVMC algorithm utilizing identical beam-geometry, MLC-positions, and monitor units and subsequently compared to clinical PB-hete plans. Results: Our results using RPC liver motion phantom validation were all compliance with MD Anderson standards. However, compared to PB-hete, average target volume encompassed by the prescribed percent isodose (Vp) was 9.1% and 8.5% less for PTV1 and PTV2 with XVMC. For the clinical liver SBRT plans, PB-hete systematically overestimated PTV dose (D95, Dmean and D10) within ±2.0% (p<0.05) compared to XVMC. Mean value of Vp was about 3.8% less with XVMC compared to PB-hete (ranged 2.9–5.7% (p<0.003)). However, mean liver dose (MLD) was 3.2% higher (p<0.003), on average, with XVMC compared to clinical PB-hete (ranged −1.0to−3.9%). OARs doses were statistically insignificant. Conclusion: Results from our XVMC dose calculations and validation study for liver SBRT indicate small-to-moderate under-dosing of the tumor volume when compared to PB-hete. Results were consistent with phantom validation and patients plans. However, Vp was less by up to 5.7% for some liver SBRT patients with XVMC–suggesting under-dosing of the target volume and overdosing of MLD by up to 3.9% occurred with PB-hete plan. These differences between PB-hete and XVMC dose calculations may be of clinical interest.

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
43
Journal Issue
6
Journal Page Range
vp.
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2016 American Association of Physicists in Medicine