Published 1994 | Version v1
Journal article

Fluoroscopic quality control

Creators

  • 1. Meditsinski Fakultet, Plovdiv (Bulgaria)

Description

Data on the quality control of 50 fluoroscopic devices are discussed. The testing methods of automatic brightness systems of the Image Intensifier are outlined. Typical function charts of such systems are presented: 1) Variation of the patient entrance skin dose rate (PESDR) vs object dimension (the most frequent PESDR is 25-80 Gy/min at 70 cm distance from the focal spot and upon patient simulation by perspex blocks 55 to 165 mm thick); 2) Variation of the exposure rate at the front of the Image Intensifier vs object dimension (1 to 30 μGy/s upon patient simulation by perspex blocks 58 to 165 mm thick). At X-rays attenuated with 1 mm copper plate and fluoroscopy with 80 kV/1 mA, the mean exposure dose rate at the front of the Image Intensifier at 90 cm distance from the focal spot varies between 0.7 and 1.4 μGy/s (5-10 mR/min). 2 figs., 11 refs. (orig.)

Additional details

Publishing Information

Journal Title
Rentgenologiya i Radiologiya
Journal Volume
33
Journal Issue
2
Journal Page Range
p. 23-26.
ISSN
0486-400X
CODEN
RENRAR

INIS

Country of Publication
Bulgaria
Country of Input or Organization
Bulgaria
INIS RN
26010629
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
BRIGHTNESS; DOSE RATES; EXPERIMENTAL DATA; FLUOROSCOPY; IMAGE PROCESSING; QUALITY ASSURANCE; QUALITY CONTROL; RADIATION DOSES
Descriptors DEC
BIOMEDICAL RADIOGRAPHY; CONTROL; DATA; DIAGNOSTIC TECHNIQUES; INFORMATION; MEDICINE; NUMERICAL DATA; OPTICAL PROPERTIES; PHYSICAL PROPERTIES