The determination of setting on X-ray equipment to get high image quality with low dosage in chest radiography using CNR ratio values
Creators
- 1. Pusat Pengkajian Sistem dan Teknologi Pengawasan Fasilitas Radiasi dan Zat Radioaktif - BAPETEN, Jl. Gajah Mada No. 8 Jakarta Pusat 10120 (Indonesia)
Description
The purpose of this assessment is to determine the best optimum setting for getting the highest image quality with lowest dose radiation received by the chest phantom using general radiography system and computed radiography (CR) system. In the study nano dot were used to measure the radiation dose on the phantom at a certain location with variation of exposure parameters. The study involved two procedures which are varies kVp with fix mA, exposure time and others parameter for procedure 1. Then followed with second procedure by varies of mA and fix others parameter after getting the best setting of kVp to get highest image quality in procedure 1. The images quality parameter in term of contrast noise ratio (CNR) and dose in sievert unit were evaluated in the study. All the results at those setting parameters were analyzed. Based on the results the best optimum setting is at 100 kVp, 200 mA and 0.1 s with dose 0.760 mSv and CNR 7.94. This is only valid for the settings selected in the study. (author)
Files
51070138.pdf
Files
(372.2 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:ed9d992861454c1401207e5f0d88f6cc
|
372.2 kB | Preview Download |
Additional details
Additional titles
- Original title (Indonesian)
- Penentuan setting penyinaran pesawat sinar-x untuk mendapatkan kualitas citra tinggi dengan dosis rendah pada radiografi dada menggunakan nilai Rasio CNR
Publishing Information
- Publisher
- Nuclear Energy Regulatory Agency
- Imprint Place
- Yogyakarta (Indonesia)
- Imprint Title
- Proceedings of the seminar on nuclear safety 2017: Development of nuclear power oversight based on public information and communication technology
- Imprint Pagination
- 389 p.
- Journal Page Range
- p. 42-47
- ISSN
- 1412-3258
- Report number
- INIS-ID--89
Conference
- Title
- The Seminar on Nuclear Safety 2017
- Original Conference Title
- Seminar Keselamatan Nuklir 2017
- Dates
- 1 Aug 2017
- Place
- Yogyakarta (Indonesia)
INIS
- Country of Publication
- Indonesia
- Country of Input or Organization
- Indonesia
- INIS RN
- 51070138
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; CONCENTRATION RATIO; MAMMARY GLANDS; RADIATION DOSES; RADIATION QUALITY; TESTING; X-RAY EQUIPMENT; X-RAY RADIOGRAPHY
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DOSES; EQUIPMENT; GLANDS; INDUSTRIAL RADIOGRAPHY; MATERIALS TESTING; MEDICINE; NONDESTRUCTIVE TESTING; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; TESTING
Optional Information
- Notes
- 4 refs.; 4 tabs.; 8 figs. Imprint:Prosiding seminar keselamatan nuklir 2017: Pengembangan pengawasan ketenaganukliran berbasis teknologi informasi dan komunikasi publik