Published December 1993 | Version v1
Journal article

A phantom for standardization and quality control in peripheral bone measurements by PQCT and DXA

  • 1. Institute for Biomedical Engineering, Federal Inst. of Technology and University of Zurich (Switzerland)
  • 2. Siemens Medical Systems, Erlangen (Germany)

Description

A phantom consisting of water- and bone-equivalent solid materials was devised to test peripheral bone densitometry systems. A circular cross section with a diameter of 60 mm was chosen, with the sides flattened by 1 cm to model the forearm. Three inserts with precisely 50, 100 and 200 mg cm-3 hydroxyapatite simulate trabecular bone and provide the basis for the linearity check of PQCT systems. Density and thickness of the simulated compact bone is selected such that the phantom can also be used to calibrate DXA and SPA machines. Quality control with CT showed a homogeneously distributed trabecular bone with an average density that deviates less than 1 mg cm-3 from linearity. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
38
Journal Issue
12
Journal Page Range
p. 1963-1970.
ISSN
0031-9155
CODEN
PHMBA7