Entrance surface doses in patients during upper gastrointestinal examination
- 1. Japanese Foundation for Cancer Research, Tokyo. Cancer Inst.
Description
The entrance surface dose distributions in patients during the upper gastrointestinal examinations in the Cancer Institute Hospital were measured in 1972 and 1976, using CaSO4:Tm foil thermoluminescent dosimeters, which were transparent for x-rays. Average values of the maximum entrance surface doses during examination with the universal gyroscopic, remotely controlled, and general x-ray television apparatus were 23, 44, and 68 R, respectively, in 1972. By improving the x-ray quality, field size, and fluoroscopic procedures, radiation exposure in fluoroscopy was reduced and, as a result, average values of the total entrance surface doses during examination with the above apparatus were reduced to 8, 22, and 28 R, respectively, in 1976. (auth.)
Additional details
Publishing Information
- Journal Title
- Nippon Igaku Hoshasen Gakkai Zasshi
- Journal Volume
- 38
- Journal Issue
- 7
- Series
- Nippon Igaku Hoshasen Gakkai Zasshi.
- Journal Page Range
- 669-681
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 10445330
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- AGE DEPENDENCE; DOSE-RESPONSE RELATIONSHIPS; FLUOROSCOPY; GASTROINTESTINAL TRACT; PATIENTS; PERSONNEL MONITORING; RADIATION DOSE DISTRIBUTIONS; SEX DEPENDENCE; SURFACES; THERMOLUMINESCENT DOSEMETERS; TIME DEPENDENCE; X RADIATION
- Descriptors DEC
- BIOMEDICAL RADIOGRAPHY; DIAGNOSTIC TECHNIQUES; DIGESTIVE SYSTEM; DOSEMETERS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MEDICINE; MONITORING; RADIATION MONITORING; RADIATIONS