Published 1982 | Version v1
Report Open

Place of HgI2 energy-dispersive x-ray detectors

Description

After a review of solid-state conduction counters, in general, and of the history of mercuric iodide, in particular, the theory of operation of solid-state energy-dispersive HgI2 detectors is dicusssed. The main factors which limit energy resolution in solid-state compound detectors are considered, including statistical fluctuations in charge generation, the window effect, trapping, inhomogeneities in the detector material, and electronic noise. Potential applications of room-temperature HgI2 x-ray detectors are listed, and general considerations are discussed for x-ray fluorescence analysis with HgI2. Directions of current investigations are given

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS., PC A03/MF A01 as DE82017651.

Files

14725205.pdf

Files (947.4 kB)

Name Size Download all
md5:74a1ed1792a815720a47d7e60fa7474a
947.4 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
41 p.
Report number
CONF-820667--6

Conference

Title
5. international workshop on HgI/sub 2/ nuclear radiation detectors.
Dates
6 - 8 Jun 1982.
Place
Jerusalem (Israel).