Published February 1984 | Version v1
Journal article

A time-of-flight hodoscope saving the number of photomultipliers

  • 1. Dipartimento di Scienze Fisiche, Universita di Cagliari, 09100 Cagliari

Description

Results of measurements of the performances of a time-of-flight hodoscope consisting of 16 slabs of plastic scintillator, 90x4x1 cm3 each, viewed by only 8 (instead of 32) photomultipliers are described. The slabs are coupled to the photomultipliers in such a way that they are considered as the elements of a square matrix of rank 4. Thus only 2x4 photomultipliers are needed to identify unambiguously each of the 16 slabs, if only a particle at a time has to be detected. By means of a very simple electronics, the authors obtained a local time-of-flight resolution of 570 ps fwhm and a spatial resolution along each slab of 5 cm fwhm. The cross-talk due to unrecognized events was less than 5%

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
31
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
146-149
ISSN
0018-9499

Conference

Title
Nuclear science symposium.
Dates
19-21 Oct 1983.
Place
San Francisco, CA (USA).

Optional Information

Secondary number(s)
CONF-831015--.