A time-of-flight hodoscope saving the number of photomultipliers
Creators
- 1. Dipartimento di Scienze Fisiche, Universita di Cagliari, 09100 Cagliari
Description
Results of measurements of the performances of a time-of-flight hodoscope consisting of 16 slabs of plastic scintillator, 90x4x1 cm3 each, viewed by only 8 (instead of 32) photomultipliers are described. The slabs are coupled to the photomultipliers in such a way that they are considered as the elements of a square matrix of rank 4. Thus only 2x4 photomultipliers are needed to identify unambiguously each of the 16 slabs, if only a particle at a time has to be detected. By means of a very simple electronics, the authors obtained a local time-of-flight resolution of 570 ps fwhm and a spatial resolution along each slab of 5 cm fwhm. The cross-talk due to unrecognized events was less than 5%
Additional details
Publishing Information
- Journal Title
- IEEE Trans. Nucl. Sci.
- Journal Volume
- 31
- Journal Issue
- 1
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 146-149
- ISSN
- 0018-9499
Conference
- Title
- Nuclear science symposium.
- Dates
- 19-21 Oct 1983.
- Place
- San Francisco, CA (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 15068316
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CHARGED PARTICLE DETECTION; CONFIGURATION; HODOSCOPES; MATRICES; NOISE; PERFORMANCE; PHOTOMULTIPLIERS; PLASTIC SCINTILLATION DETECTOR; SPATIAL RESOLUTION; TIME RESOLUTION; TIME-OF-FLIGHT METHOD
- Descriptors DEC
- DETECTION; MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; TIMING PROPERTIES