Published 2009
| Version v1
Book
Evaluating the influence of the aluminum purity at mammographic HVL procedures
- 1. Fisica Medica Servicos e Consultoria Tecnica S/C, Sao Paulo, SP (Brazil)
- 2. Universidade de Sao Paulo (USP), SP (Brazil). Inst. de Fisica
- 3. Universidade de Sao Paulo (IEE/USP), SP (Brazil). Inst. de Eletrotecnica e Energia
Description
This paper intends to evaluate the influence of the Aluminum's purity over the mammographic HVL procedures. It was measured the HVL of a GE Senographe 700 T using 7 groups of commercial filters, and its purity and thickness were determined. As the results shows, the first HVL inherent procedure's uncertainty is about 3%; the best local prototype's deviation over the worst commercial device is about 5%; and, according to the procedure's uncertainty, as to the commercial device's response variability, it may be possible using local Al filters, although it is necessary to extend this study over more equipment in order to set up a correction factor, which was found to be about 5% to 8% over Al filters about 99% purity, on this study. (author)
Additional details
Additional titles
- Original title (Portuguese)
- Avaliacao da influencia do teor de pureza do aluminio nos resultados do teste da camada semi-redutora em mamografia
Publishing Information
- Publisher
- ABFM
- Imprint Place
- Sao Paulo (Brazil)
- ISBN
- 978-85-85205-94-2
- Imprint Title
- Proceedings of the 14. Brazilian congress on medical physics; 4. Symposium on instrumentation and medical imaging
- Imprint Pagination
- [vp.]
- Journal Page Range
- [4 p.]
Conference
- Title
- 14. Brazilian congress on medical physics; 4. Symposium on instrumentation and medical imaging
- Original Conference Title
- 14. Congresso brasileiro de fisica medica
- Dates
- 8-12 Jul 2009
- Place
- Sao Paulo, SP (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 41113794
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ABSORPTION; ALUMINIUM; BIOLOGICAL SHIELDING; BIOPHYSICS; DIAGNOSIS; DIAGNOSTIC TECHNIQUES; DOSIMETRY; EXTERNAL IRRADIATION; HALF-THICKNESS; QUALITY ASSURANCE; QUALITY CONTROL; RADIATION HAZARDS; RADIATION LENGTH; RADIATION MONITORING; RADIATION PROTECTION; RADIATION SOURCES; RADIOLOGY; RADIOPROTECTIVE SUBSTANCES; RELIABILITY; SHIELDING; THICKNESS
- Descriptors DEC
- CONTROL; DIMENSIONS; DRUGS; ELEMENTS; HAZARDS; HEALTH HAZARDS; IRRADIATION; LENGTH; MEDICINE; METALS; MONITORING; NUCLEAR MEDICINE; PHYSICAL PROPERTIES; PHYSICS; RESPONSE MODIFYING FACTORS; SHIELDING; SORPTION
Optional Information
- Notes
- Imprint:Anais do 14. Congresso brasileiro de fisica medica; 4. Simposio de instrumentacao e imagens medicas