Comparison of quantitative image quality of cardiac computed tomography between raw-data-based and model-based iterative reconstruction algorithms with an emphasis on image sharpness

  • 1. Asan Medical Center, Department of Radiology and Research Institute of Radiology, University of Ulsan College of Medicine, Songpa-gu, Seoul (Korea, Republic of)

Description

Image sharpness is commonly degraded on cardiac CT images reconstructed using iterative reconstruction algorithms. To compare the image quality of cardiac CT between raw-data-based and model-based iterative reconstruction algorithms developed by the same CT vendor in children and young adults with congenital heart disease. In 29 patients with congenital heart disease, we reconstructed 39 cardiac CT datasets using raw-data-based and model-based iterative reconstruction algorithms. We performed quantitative analysis of image sharpness using distance25â75% and angle25â75% on a line density profile across an edge of the descending thoracic aorta in addition to CT attenuation, image noise, signal-to-noise ratio and contrast-to-noise ratio. We compared these quantitative image-quality measures between the two algorithms. CT attenuation did not show significant differences between the algorithms (P>0.05) except in the aorta. Image noise was small but significantly higher in the model-based algorithm than in the raw-data-based algorithm (4.8±2.3 Hounsfield units [HU] vs. 4.7±2.1 HU, P<0.014). Signal-to-noise ratio (110.2±50.9 vs. 108.4±50.4, P=0.050) and contrast-to-noise ratio (91.0±45.7 vs. 89.6±45.1, P=0.063) showed marginal significance between the two algorithms. The model-based algorithm showed a significantly smaller distance25â75% (1.4±0.4 mm vs. 1.6±0.3 mm, P<0.001) and a significantly higher angle25â75% (77.0±4.3° vs. 74.1±5.7°, P<0.001) than the raw-data-based algorithm. Compared with the raw-data-based algorithm, the model-based iterative reconstruction algorithm demonstrated better image sharpness and higher image noise on cardiac CT in patients with congenital heart disease.

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00247-020-04741-x

Additional details

Identifiers

Publishing Information

Journal Title
Pediatric Radiology
Journal Volume
50
Journal Issue
11
Journal Page Range
p. 1570-1578
ISSN
0301-0449
CODEN
PDRYA5