Published February 2013 | Version v1
Journal article

Comparison of activity measurements in nuclear medicine services in the Czech Republic

  • 1. Czech Metrology Institute, Prague (Czech Republic)

Description

Radionuclide calibrator used in nuclear medicine is the radiation detection equipment for measuring radiopharmaceutical activity. This instrument consists of an ionization chamber with an electrometer. This paper examines the variation in the amounts of radioactivity that is associated with dose calibration. The amount of radioactivity in each sample was measured by CMI and compared with values obtained from radionuclide calibrators over a year period. Measurements of serum bottles were within 10 % for 18F and 90Y, 10-15 % for 131I and 51Cr, 10-20 % for 99mTc, 15-20 % for 67Ga, 10-25 % for 111In, 15-30 % for 201Tl and 15-45 % for 123I. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
295
Journal Issue
2
Journal Page Range
p. 1107-1111
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
9 refs.