Patient-specific stopping power calibration for proton therapy planning based on single-detector proton radiography
- 1. Department of Medical Physics and Bioengineering, University College London, London (United Kingdom)
- 2. Department of Radiation Oncology, Massachusetts General Hospital, Boston, MA (United States)
- 3. Ion Beam Applications (IBA), 3 Chemin du Cyclotron, Louvain la Neuve, B-1348 (Belgium)
Description
A simple robust optimizer has been developed that can produce patient-specific calibration curves to convert x-ray computed tomography (CT) numbers to relative stopping powers (HU-RSPs) for proton therapy treatment planning. The difference between a digitally reconstructed radiograph water-equivalent path length (DRRWEPL) map through the x-ray CT dataset and a proton radiograph (set as the ground truth) is minimized by optimizing the HU-RSP calibration curve. The function of the optimizer is validated with synthetic datasets that contain no noise and its robustness is shown against CT noise. Application of the procedure is then demonstrated on a plastic and a real tissue phantom, with proton radiographs produced using a single detector. The mean errors using generic/optimized calibration curves between the DRRWEPL map and the proton radiograph were 1.8/0.4% for a plastic phantom and −2.1/ − 0.2% for a real tissue phantom. It was then demonstrated that these optimized calibration curves offer a better prediction of the water equivalent path length at a therapeutic depth. We believe that these promising results are suggestive that a single proton radiograph could be used to generate a patient-specific calibration curve as part of the current proton treatment planning workflow. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/60/5/1901Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 60
- Journal Issue
- 5
- Journal Page Range
- p. 1901-1917
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47004668
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CALIBRATION; COMPUTERIZED TOMOGRAPHY; DATASETS; DEPTH; ERRORS; FORECASTING; IMAGES; LENGTH; NOISE; OPTIMIZATION; PATIENTS; PHANTOMS; PLANNING; PROTON BEAMS; PROTON RADIOGRAPHY; RADIOTHERAPY; STOPPING POWER; X RADIATION
- Descriptors DEC
- BEAMS; DIAGNOSTIC TECHNIQUES; DIMENSIONS; DOCUMENT TYPES; ELECTROMAGNETIC RADIATION; INDUSTRIAL RADIOGRAPHY; IONIZING RADIATIONS; MATERIALS TESTING; MEDICINE; MOCKUP; NONDESTRUCTIVE TESTING; NUCLEAR MEDICINE; NUCLEON BEAMS; PARTICLE BEAMS; RADIATIONS; RADIOLOGY; STRUCTURAL MODELS; TESTING; THERAPY; TOMOGRAPHY