Published October 1997 | Version v1
Journal article

New approach to improving the reliability of X-ray fluorescence analysis

  • 1. Inst. Gornogo Dela im. A.A. Skochinskogo RAN, Lyubertsy (Russian Federation)

Description

It is shown that the probability of a mistake in the recognition of compositions close to each other in X-ray fluorescence analysis can be decreased by several tens of orders of magnitude if sets of signals from the X-ray spectrometer are selected to be specially correlated to each other. It is proved that the reliability of control can be sharply improved only with the use of sets of signals for which the longer axes of ellipsoids of equal probability densities are nearly parallel to each other and the ellipoids themselves exhibit a high eccentricity. Informativeness tests are analytically deduced for sets of two, three and n signals with different covariance matrices. The root-mean-square deviation of the measurement error is determined from the informativeness test calculated for the given set of closely correlated signals

Additional details

Additional titles

Original title (Russian)
Новый подход к повышению точности рентгенофлуоресцентного анализа

Publishing Information

Journal Title
Zhurnal Analiticheskoj Khimii
Journal Volume
52
Journal Issue
10
Journal Page Range
p. 1030-1035.
ISSN
0044-4502
CODEN
ZAKHA8

INIS

Country of Publication
Russian Federation
Country of Input or Organization
Russian Federation
INIS RN
29022958
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Descriptors DEI
ACCURACY; CALIBRATION; CHEMICAL COMPOSITION; ERRORS; MATHEMATICAL MODELS; X-RAY FLUORESCENCE ANALYSIS; X-RAY FLUORESCENCE ANALYZERS
Descriptors DEC
CHEMICAL ANALYSIS; NONDESTRUCTIVE ANALYSIS; X-RAY EMISSION ANALYSIS

Optional Information

Notes
14 refs.; 2 figs.