Published November 7, 2005 | Version v1
Journal article

Polarity effect of the thimble-type ionization chamber at a low dose rate

  • 1. Korea Atomic Energy Research Institute, Daejeon 305-600 (Korea, Republic of)
  • 2. Environmental Engineering, Yonsei University, Won-Ju 220-710 (Korea, Republic of)
  • 3. iTRS, Hanyang University, Hangdang-dong, Seoul 133-791 (Korea, Republic of)

Description

It is known that the current collected from an ionization chamber exposed to a constant radiation intensity changes in magnitude when the polarity of the collecting potential is reversed. It is called the polarity effect of the ionization chamber. There are many possible causes that induce the polarity effect and one of them can be a field distortion due to a potential difference between the guard electrode and the collector. We studied how much the polarity effect depends on the design of the electrodes in the thimble-type ionization chamber. Two thimble-type ionization chambers, which had different electrode structures, were designed and fabricated at KAERI. We calculated the field distortions due to the potential difference between the guard electrode and the collector for the two ionization chambers. MAXWELL and Garfield were employed to calculate the electron drift lines inside the chamber. The polarity effects of the two ionization chambers were measured, and they were consistent with the field calculation. We could conclude that the polarity effect is mostly induced from the field distortion due to the potential difference between the guard electrode and the collector in our experiment and it depends significantly on the design of the electrodes

Availability note (English)

Available online at http://stacks.iop.org/0031-9155/50/4995/pmb5_21_003.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
50
Journal Issue
21
Journal Page Range
p. 4995-5003
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
37048613
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
DESIGN; DOSE RATES; ELECTRODES; ELECTRON DRIFT; IONIZATION CHAMBERS; KAERI; SECURITY PERSONNEL
Descriptors DEC
KOREAN ORGANIZATIONS; MEASURING INSTRUMENTS; NATIONAL ORGANIZATIONS; PERSONNEL; RADIATION DETECTORS