Published 1992 | Version v1
Journal article

The role of phantom parameters on the response of the AEOI Neutriran Albedo Neutron Personnel Dosemeter

  • 1. Atomic Energy Organization of Iran, Teheran (Iran, Islamic Republic of)

Description

The response of the AEOI Neutriran Albedo Neutron Personnel Dosemeter (NANPD) which can also be used for other albedo dosemeter types was determined on 18 different phantom configurations. The effects of type, geometry, material, thickness, dosemeter-to-phantom angle in particular with the presence of legs were investigated using a Pu-Be neutron source. It was concluded that the slab phantoms (single or double) and circular and elliptical cylinder phantoms seemed to provide a better response, whereas the ICRU sphere geometry does not seem to be appropriate for the calibration of albedo dosemeters. It is interesting to note that the presence of legs maintains the constancy of the response in a situation when a radiation worker bends down during work. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
44
Journal Issue
1-4
Journal Page Range
p. 281-285.
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
7. Symposium on neutron dosimetry.
Dates
14-18 Oct 1991.
Place
Berlin (Germany).

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
24038352
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALBEDO-NEUTRON DOSEMETERS; GEOMETRY; LEGS; NEUTRON DOSIMETRY; PERFORMANCE; PERSONNEL DOSIMETRY; PHANTOMS; SIZE; THICKNESS
Descriptors DEC
BODY; BODY AREAS; DIMENSIONS; DOSEMETERS; DOSIMETRY; LIMBS; MATHEMATICS; MEASURING INSTRUMENTS; MOCKUP; STRUCTURAL MODELS