Published 1997
| Version v1
Journal article
Realistic Monte Carlo simulation of the INOR Therapy Unit and electron contamination calculation for a board 60Co Beam
Creators
- 1. Laboratory for Nuclear Analysis, Cuba (Cuba). Center of Applied Studies for Nuclear Development
Description
The therapy beam from the 60Co unit used at INOR was modeled using the EGS Monte Carlo radiation transport system. The rectangular geometry of the collimating system is fully simulated and the effect of considering the outer collimator a set of stepped collimators rather than a collimator with a smooth surface was estimated. Variance reduction techniques are widely used to increase the efficiency of our calculations. Even for the dose at very shallow depths (0.5mm) the agreement of our calculations with the experimental results is very good. The shift of peak dose to about 1mm due to electron contamination is observed. (author)
Availability note (English)
Also available from Biblioteca Luis Demetrio Tinoco, Universidad de Costa RicaAdditional details
Publishing Information
- Journal Title
- Tecnologia en Marcha
- Journal Volume
- 13
- Journal Issue
- special issue
- Journal Page Range
- p. 134-139
- ISSN
- 0379-3982
INIS
- Country of Publication
- Costa Rica
- Country of Input or Organization
- Costa Rica
- INIS RN
- 31040044
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BEAMS; CALCULATION METHODS; COLLIMATORS; DEPTH DOSE DISTRIBUTIONS; EFFICIENCY; GEOMETRY; MONTE CARLO METHOD; RADIATION DOSES; REDUCTION
- Descriptors DEC
- CALCULATION METHODS; CHEMICAL REACTIONS; DOSES; MATHEMATICS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS
Optional Information
- Notes
- Refs., figs.