Image noise-based dose adaptation in dynamic volume CT of the heart: dose and image quality optimisation in comparison with BMI-based dose adaptation
Creators
- 1. Queen's University, School of Medicine, Kingston, ON (Canada)
- 2. University of Toronto, Division of Cardiothoracic Imaging, Department of Medical Imaging, Toronto General Hospital, Toronto, ON (Canada)
- 3. Toshiba Medical Systems Europe BV, Zoetermeer (Netherlands)
Description
To compare the image quality and radiation dose using image-noise (IN)-based determination of X-ray tube settings compared with a body mass index (BMI)-based protocol during CT coronary angiography (CTCA). Two hundred consecutive patients referred for CTCA to our institution were divided into two groups: BMI-based, 100 patients had CTCA with the X-ray tube current adjusted to the patient's BMI while maintaining a fixed tube potential of 120 kV; IN-based, 100 patients underwent imaging with the X-ray tube current and voltage adjusted to the IN measured within the mid-left ventricle on a pre-acquisition trans-axial image. Two independent cardiac radiologists performed blinded image quality assessment with quantification of the IN and signal-to-noise ratio (SNR) from the mid-LV and qualitative assessment using a three-point score. Radiation dose (CTDI and DLP) was recorded from the console. Results showed: IN (HU): BMI-based, 30.1 ± 9.9; IN-based, 33.1 ± 6.7; 32 % variation reduction (P = 0.001); SNR: BMI-based, 18.6 ± 7.1; IN-based, 15.4 ± 3.7; 48 % variation reduction (P < 0.0001). Visual scores: BMI-based, 2.3 ± 0.6; IN-based, 2.2 ± 0.5 (P = 0.54). Radiation dose: CTDI (mGy), BMI-based, 22.68 ± 8.9; IN-based, 17.16 ± 7.6; 24.3 % reduction (P < 0.001); DLP (mGy.cm), BMI-based, 309.3 ± 127.5; IN-based, 230.6 ± 105.5; 25.4 % reduction (P < 0.001). Image-noise-based stratification of X-ray tube parameters for CTCA results in 32 % improvement in image quality and 25 % reduction in radiation dose compared with a BMI-based protocol. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-013-2980-1Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 24
- Journal Issue
- 1
- Journal Page Range
- p. 86-94
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 45040275
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; CORONARIES; DATA ACQUISITION; HEART; OPTIMIZATION; PATIENTS; RADIATION DOSES; SIGNAL-TO-NOISE RATIO; SIZE; X-RAY TUBES
- Descriptors DEC
- ARTERIES; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DOSES; ELECTRON TUBES; EQUIPMENT; EVALUATION; ORGANS; TOMOGRAPHY; X-RAY EQUIPMENT