Published April 1991
| Version v1
Journal article
A bi4π counting method for activity measurement
Description
The bi4π counting method as a new fundamental method of activity measurement is introduced. Suitable for more radionuclides, the method can achieve an accuracy of 0.03-0.6% (1σ) in principle. The correctness of this method is verified by the measurements of 60Co sources, and it is found readily to reach an accuracy of 0.2%. The results are compared with those obtained by the 4πβ-γ coincident method, and both are in agreement within the measurement error. Finally, the range of application for this method is also discussed
Additional details
Publishing Information
- Journal Title
- Nuclear Techniques
- Journal Volume
- 14
- Journal Issue
- 4
- Series
- Nucl. Tech.
- Journal Page Range
- 228-233
- ISSN
- 0253-3219
- CODEN
- NUTED
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 23075460
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- ACTIVITY LEVELS; COBALT 60; COMPARATIVE EVALUATIONS; CORRECTIONS; DATA COVARIANCES; DEAD TIME; EFFICIENCY; EXTRAPOLATION; FOUR-PI COUNTING; PLASTIC SCINTILLATION DETECTOR; PROPORTIONAL COUNTERS; SELF-ABSORPTION
- Descriptors DEC
- ABSORPTION; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COBALT ISOTOPES; COUNTING TECHNIQUES; EVALUATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; NUMERICAL SOLUTION; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SORPTION; TIMING PROPERTIES; YEARS LIVING RADIOISOTOPES