Published August 1, 1992 | Version v1
Journal article

Four channel V to F converter with frequency counters

  • 1. Brookhaven National Lab., Upton, NY (United States)

Description

A high linearity four channel voltage to frequency converter has been developed for general purpose beamline data collection. The full scale frequency output for each channel is 1 or 2 MHz. Inputs are differential and input voltage ranges are independently selectable to various unipolar and bipolar ranges. Each channel is also equipped with a 7 1/2 digit frequency counter with independently selectable gate times. The entire unit is packaged in a single with NIM module. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A
Journal Volume
319
Journal Issue
1-3
Journal Page Range
p. 414-416.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
7. national conference and exhibition on synchrotron radiation instrumentation (SRI-7).
Dates
28-31 Oct 1991.
Place
Baton Rouge, LA (United States).

INIS

Country of Publication
Netherlands
Country of Input or Organization
Netherlands
INIS RN
24002597
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BACKGROUND NOISE; BEAM TRANSPORT; DATA ACQUISITION SYSTEMS; INTEGRATED CIRCUITS; MODULAR STRUCTURES; NUCLEAR INSTRUMENT MODULES; PULSE CONVERTERS
Descriptors DEC
ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; EQUIPMENT; NOISE