Description

We describe the design of a wide temperature range (300 mK-300 K) atomic force microscope/magnetic force microscope with a self-aligned fibre-cantilever mechanism. An alignment chip with alignment groves and a special mechanical design are used to eliminate tedious and time consuming fibre-cantilever alignment procedure for the entire temperature range. A low noise, Michelson fibre interferometer was integrated into the system for measuring deflection of the cantilever. The spectral noise density of the system was measured to be ∼12 fm/√Hz at 4.2 K at 3 mW incident optical power. Abrikosov vortices in BSCCO(2212) single crystal sample and a high density hard disk sample were imaged at 10 nm resolution to demonstrate the performance of the system

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
85
Journal Issue
10
Journal Page Range
p. 103705-103705.7
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2014 AIP Publishing LLC