Published October 15, 2015 | Version v1
Journal article

An interface for the direct coupling of small liquid samples to AMS

  • 1. Center for Accelerator Mass Spectrometry, Lawrence Livermore National Laboratory, 7000 East Ave, Livermore, CA 94551 (United States)
  • 2. Biology and Biotechnology Division, Lawrence Livermore National Laboratory, 7000 East Ave, Livermore, CA 94551 (United States)

Description

We describe the moving wire interface attached to the 1-MV AMS system at LLNL's Center for Accelerator Mass Spectrometry for the analysis of nonvolatile liquid samples as either discrete drops or from the direct output of biochemical separatory instrumentation, such as high-performance liquid chromatography (HPLC). Discrete samples containing at least a few 10 s of nanograms of carbon and as little as 50 zmol 14C can be measured with a 3–5% precision in a few minutes. The dynamic range of our system spans approximately 3 orders in magnitude. Sample to sample memory is minimized by the use of fresh targets for each discrete sample or by minimizing the amount of carbon present in a peak generated by an HPLC containing a significant amount of 14C. Liquid sample AMS provides a new technology to expand our biomedical AMS program by enabling the capability to measure low-level biochemicals in extremely small samples that would otherwise be inaccessible.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nimb.2015.05.018

Additional details

Identifiers

DOI
10.1016/j.nimb.2015.05.018;
PII
S0168-583X(15)00489-9;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
Journal Volume
361
Journal Page Range
p. 173-177
ISSN
0168-583X
CODEN
NIMBEU

Conference

Title
13. accelerator mass spectrometry conference
Acronym
AMS-13
Dates
24-29 Aug 2014
Place
Aix en Provence (France)

Optional Information

Copyright
Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.