Report Open

The need for a quality control program in emission tomography

Description

It is particularly important for a single photon tomographic system (SPECT) that the measurement conditions be-well-defined. Even then, traps exist. It has been observed, in comparing two ''identical'' systems that, for the same resolution, sensitivity, uniformity etc, differences of almost an order in the apparent contrast were observed. This effect was a result of a slightly different energy window setting, and caused by the effect of scatter increasing the tail effect. The definition of a set of standard phantoms and measurement protocols should be a high priority for one of the international standards agencies. Without such a programme, it is possible that some centres will use SPECT systems without them being properly set-up, and, as a result, obtain potential very high clinical false positive rates

Availability note (English)

MF available from INIS under the Report Number.

Files

13678461.pdf

Files (255.4 kB)

Name Size Download all
md5:49ce33b460ee0bc50781b9bd1feb410e
255.4 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
4 p.
Report number
CEA-CONF--6044

Conference

Title
19. International meeting of the Society of nuclear medecine.
Dates
8 - 11 Sep 1981.
Place
Bern, Switzerland.

INIS

Country of Publication
France
Country of Input or Organization
France
INIS RN
13678461
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
QUALITY CONTROL; RESOLUTION; SENSITIVITY; SINGLE PHOTON ECT
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; CONTROL; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; TOMOGRAPHY