Published November 2014 | Version v1
Journal article

Radiation doses in adult computed tomography practice in Serbia: initial results

  • 1. School of Electrical Engineering, University of Belgrade, Belgrade (Serbia)
  • 2. Vinca Institute of Nuclear Sciences, University of Belgrade, Belgrade (Serbia)
  • 3. Clinical Centre of Vojvodina, Novi Sad (Serbia)

Description

This work presents initial data on radiation doses in adult computed tomography (CT) in Serbia. Data were collected in terms of CT dose index (CTDIvol) and dose length product (DLP) values for head, chest and abdomen examination. The range of CTDIvol values was found to be 53-98, 11-34 and 8.5-227 mGy whereas for DLP was 803-1066, 350-845 and 1066-3078 mGy cm-1 for head, chest and abdomen examination, respectively. Except for abdomen on one CT unit, all estimated values were in line with the reported data. This work also presents simple method on how to reduce radiation doses when scanning head. Using axial (step-and-shot) instead of helical mode and decreasing tube current-time product leads to significant dose reduction. CTDIvol was decreased by 20 % whereas DLP was reduced for a factor 2. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncu245

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
162
Journal Issue
1-2
Journal Page Range
p. 135-138
ISSN
0144-8420

Conference

Title
2. International Conference on Radiation and Dosimetry in Various Fields of Research; 2. East European Radon Symposium - SEERAS
Acronym
RAD 2014
Dates
27-30 May 2014
Place
Nis (Serbia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
46037058
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CHEST; COMPUTERIZED TOMOGRAPHY; RADIATION DOSES; RADIATION PROTECTION; SERBIA
Descriptors DEC
BODY; DEVELOPING COUNTRIES; DIAGNOSTIC TECHNIQUES; DOSES; EASTERN EUROPE; EUROPE; TOMOGRAPHY

Optional Information

Notes
21 refs.