Published October 10, 1986
| Version v1
Journal article
Gas-chromatographic analysis of high-purity helium using a helium detector
- 1. Institute of Chemistry, Gor'kii, USSR
Description
The limits of gas-chromatographic detection of neon, hydrogen, argon, nitrogen, krypton, and methane in helium have been determined using a helium ionization detector under saturation current conditions. The detection limits are restricted by the gas permeability of the detector Teflon body and the injection system. The dependence of extraction of impurity gases by cryogenic adsorption enrichment on their contents and enrichment time has been examined. the relative detection limit can be lowered by preconcentration of 3 x 10-5% for neon and to 4 x 10-7 to 2 x 10-8% for other gases
Additional details
Publishing Information
- Journal Title
- J. Anal. Chem. USSR (Engl. Transl.)
- Journal Volume
- 41
- Journal Issue
- 4
- Series
- J. Anal. Chem. USSR (Engl. Transl.).
- Journal Page Range
- 552-557
- ISSN
- 0021-8766
- CODEN
- JACUA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 19039560
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Resource subtype / Literary indicator
- Translation
- Descriptors DEI
- ADSORPTION; ARGON; CALIBRATION; CRYOGENICS; EFFICIENCY; GAS CHROMATOGRAPHY; HELIUM; HYDROGEN; IMPURITIES; IONIZATION; KRYPTON; LIMITING VALUES; METHANE; NEON; NITROGEN; OXYGEN; PERMEABILITY; PRODUCTION; PURIFICATION; SEPARATION EQUIPMENT; TEFLON
- Descriptors DEC
- ALKANES; CHROMATOGRAPHY; ELEMENTS; EQUIPMENT; HYDROCARBONS; NONMETALS; ORGANIC COMPOUNDS; ORGANIC FLUORINE COMPOUNDS; ORGANIC HALOGEN COMPOUNDS; ORGANIC POLYMERS; POLYETHYLENES; POLYMERS; POLYOLEFINS; RARE GASES; SEPARATION PROCESSES
Optional Information
- Notes
- Translated from Zh. Anal. Khim.; 41: No. 4, 696-701(Apr 1986).