Published October 10, 1986 | Version v1
Journal article

Gas-chromatographic analysis of high-purity helium using a helium detector

  • 1. Institute of Chemistry, Gor'kii, USSR

Description

The limits of gas-chromatographic detection of neon, hydrogen, argon, nitrogen, krypton, and methane in helium have been determined using a helium ionization detector under saturation current conditions. The detection limits are restricted by the gas permeability of the detector Teflon body and the injection system. The dependence of extraction of impurity gases by cryogenic adsorption enrichment on their contents and enrichment time has been examined. the relative detection limit can be lowered by preconcentration of 3 x 10-5% for neon and to 4 x 10-7 to 2 x 10-8% for other gases

Additional details

Publishing Information

Journal Title
J. Anal. Chem. USSR (Engl. Transl.)
Journal Volume
41
Journal Issue
4
Series
J. Anal. Chem. USSR (Engl. Transl.).
Journal Page Range
552-557
ISSN
0021-8766
CODEN
JACUA

Optional Information

Notes
Translated from Zh. Anal. Khim.; 41: No. 4, 696-701(Apr 1986).