In calculations for the Leksell Gamma Knife, GammaPlan employs a tissue equivalent material without the presence of a skull bone, while dosimetry work is based on a polystyrene phantom. The compatibility of these dose distributions is uncertain. The Monte Carlo technique was employed to determine the radial dose distributions from a single 14 mm collimator helmet in 160 mm diam phantoms with different materials. The materials studied were polystyrene, perspex, water, and water with skull bone. Results showed no significant differences among the radial doses in different phantom materials for the 14 mm collimator helmet. The Monte Carlo simulation was repeated with the inclusion of all 201 sources. Again, no significant differences were observed