Published January 1, 2002 | Version v1
Journal article

Improved response function calculations for scintillation detectors using an extended version of the MCNP code

Description

The analysis of (e,e'n) experiments at the Darmstadt superconducting electron linear accelerator S-DALINAC required the calculation of neutron response functions for the NE213 liquid scintillation detectors used. In an open geometry, these response functions can be obtained using the Monte Carlo codes NRESP7 and NEFF7. However, for more complex geometries, an extended version of the Monte Carlo code MCNP exists. This extended version of the MCNP code was improved upon by adding individual light-output functions for charged particles. In addition, more than one volume can be defined as a scintillator, thus allowing the simultaneous calculation of the response for multiple detector setups. With the implementation of 12C(n,n'3α) reactions, all relevant reactions for neutron energies En<20 MeV are now taken into consideration. The results of these calculations were compared to experimental data using monoenergetic neutrons in an open geometry and a 252Cf neutron source in the complex Darmstadt setup, where in both cases excellent agreement was found

Additional details

Identifiers

PII
S016890020101422X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
476
Journal Issue
1-2
Journal Page Range
p. 155-159
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.